Page 82 - ITECO Catalogue 2024
P. 82
DATALOGGER All our ovens are equipped with a data logger for traceability and you can print/save graphs of
temperature (for sahara ovens) and temperature/humidity (for Sahara DRY ovens)
To analyze the logged data and to set the thermal programmer is required the 8107.215 op on
which inlude the PidManager so ware and USB specific cable.
The so ware is compa ble with Windows XP/../11.
Sample rate: 1, 5, 10, 15,30 sec 1, 5, 10, 15, 30 min
Dura on (T) Sahara from 6 hour(1sec) to 450 day(30min)
Dura on (T+RH%) Sahara DRY: from 3 hour(1sec) to 225 day(30min)
8107.215 PIDManager so ware and specifi c USB cable to set the thermal
programs and to download and view datalogger records
Thermal Program :
1 SETPOINT 40°C / gradient MAX.
2 WAIT 20min @40°C
3 SETPOINT 100°C / GRADIENT<2°C/min
4 WAIT 20min @100°C
5 SETPOINT 150°C / GRADIENT<1°C/min
6 WAIT 20min @150°C
Sahara DRY 80L 40°C < 5% RH Sahara 80L Thermal Program